After liquid-liquid extraction using methyl tert-butyl ether, analysis was performed on an Agilent Technologies 1290 Infinity LC system coupled with an Agilent Technologies 6460 Triple Quadrupole MS. Separation with a C18 column and gradient elution at 0.5 mL/min resulted in a 6-min run-time. Detection was performed in dynamic MRM, monitoring 3 ion transitions per compound. Isotope labeled internal standards were used for every compound, except for bromperidol and levosulpiride.
Mean recovery was 86.8%. Matrix effects were 鈭?#xA0;18.4 to + 9.1%. Accuracy ranged between 91.3 and 107.0% at low, medium and high concentrations and between 76.2 and 113.9% at LLOQ. Within-run precision was < 15% (CV), except for asenapine and hydroxy-iloperidone. Between-run precision was aberrant only for 7-hydroxy-N-desalkylquetiapine, asenapine and reduced haloperidol. No interferences were found. No problems of instability were observed, even for olanzapine. The method was successfully applied on patient samples.
The liquid-liquid extraction and UHPLC-MS/MS technique allows robust target screening and quantification of 23 antipsychotics and metabolites.