A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery
详细信息    查看全文
文摘
To investigate agreement in intraocular pressure (IOP) measurements among Icare rebound tonometry, Tonopen tonometry, and Goldmann applanation tonometry following vitreoretinal surgery.

Design

Reliability analysis of tonometers.

Methods

Fifty-eyes of 50 adults undergoing vitreoretinal surgery were enrolled. IOP was measured on first postoperative day using Icare (Tiolat, Helsinki, Finland), followed by Tonopen (Reichert, Depew, New York, USA) and Goldmann (Haag-Streit USA, Mason, Ohio, USA) in randomized order. Intraclass correlation coefficients (ICC) and Bland-Altman plots were calculated for all subcategories.

Results

Icare successfully measured IOP in all eyes, while Goldmann was unmeasurable in 6 eyes (12%) and Tonopen in 1 eye (2%). Mean IOP by Icare, Tonopen, and Goldmann was 15.9 ± 8.9, 16.9 ± 6.2, and 16.0 ± 7.3 mm Hg, respectively (P = .76). Type of intraocular tamponade, status of lens, status of cornea, gauge of instrumentation, and history of prior vitrectomy did not result in significant differences among the 3 tonometers. ICC was excellent (>0.75) in all subgroups, except at IOP <10 and ≥23 mm Hg (based on Icare). In eyes with IOP <10 mm Hg, Icare underestimated IOP (mm Hg; P = .01) compared to Goldmann (2.0 ± 2.1) and Tonopen (3.5 ± 2.4), whereas at IOP ≥23 mm Hg Icare was overestimated (P = .01) compared to Goldmann (3.77 ± 3.49) and Tonopen (4.97 ± 3.33). Overall, differences in IOP were ≤3 mm Hg in 58% of eyes for Icare-Tonopen, 72% for Tonopen-Goldmann, and 62% for Icare-Goldmann.

Conclusion

IOP measurements using Icare rebound tonometry, Tonopen, and Goldmann tonometry are in excellent agreement following vitreoretinal surgery. However, Icare overestimates at IOP ≥23 and underestimates at IOP <10 mm Hg.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700