In this work, a transmit-only receive-only dual resonator system was designed and developed to cover the whole human spine using 23Na MRI and increase the receive sensitivity. The resonator system consisted of an already presented 23Na whole-body resonator and a newly developed 5-channel receive-only phased-array. The resonator system was first validated using bench top and phantom measurements. A threefold SNR improvement at the depth of the spine (∼7 cm) over the whole-body resonator was achieved using the spine array. 23Na MR measurements of the human spine using the transmit-only receive-only resonator system were performed on a healthy volunteer within an acquisition time of 10 minutes. A density adapted 3D radial sequence was chosen with 6 mm isotropic resolution, 49 ms repetition time and a short echo time of 540 μs. Furthermore, it was possible to quantify the tissue sodium concentration in the intervertebral discs in the lumbar region (120 ms repetition time) using this setup.