An in-vivo evaluation of a MEMS drug delivery device using Kunming mice model
详细信息    查看全文
文摘
The use of MEMS implantable drug delivery pump device enables one to program the desired drug delivery profile in the device for individualized medicine treatment to patients. In this study, a MEMS drug delivery device is prepared and employed for in vivo applications. 12 devices are implanted subcutaneously into Kunming mice for evaluating their long term biocompatibility and drug-delivery efficiency in vivo. All the mice survived after device implantation surgery procedures. Histological analysis result reveals a normal wound healing progression within the tissues-to-device contact areas. Serum analysis shows that all measured factors are within normal ranges and do not indicate any adverse responses associated with the implanted device. Phenylephrine formulation is chosen and delivered to the abdominal cavity of the mice by using either the implanted MEMS device (experimental group) or the syringe injection method (control group). Both groups show that they are able to precisely control and manipulate the increment rate of blood pressure in the small animals. Our result strongly suggests that the developed refillable implantable MEMS devices will serve as a viable option for future individualized medicine applications such as glaucoma, HIV-dementia and diabetes therapy.