摘要
目的对保山市人民医院检验科贝克曼库尔特LH750血细胞分析仪进行性能验证。方法参照WS/T 406-2012要求,通过本底计数、携带污染、批内精密度、批间精密度、线性范围、准确度、正确度试验进行性能验证。结果贝克曼库尔特LH750血细胞分析仪本底计数,携带污染率,白细胞计数(WBC)、红细胞计数(RBC)、血红蛋白(Hb)、血细胞比容(HCT)、红细胞平均体积(MCV)、红细胞平均血红蛋白量(MCH)、红细胞血红蛋白浓度(MCHC)的批内、批间精密度以及准确度、正确度、线性均符合WS/T 406-2012要求。结论贝克曼库尔特LH750血细胞分析仪性能良好,测定结果准确可靠,能满足临床检测要求的标准。
Objective To perform the performance verification of the Beckman Coulter LH750 blood cell analyzer in the Baoshan Municipal People′s Hospital.Methods According to the requirements of WS/T 406-2012,the performance verification was carried out through the background counting,carry-over rate,intrabatch and inter-batch precision,linear range,accuracy and crectness tests.Results The background count of Beckman Coulter LH750 blood cell analyzer,carry over rate,intra-batch and inter-batch precision of WBC,RBC,Hb,HCT,MCV,MCH,MCHC,accuracy,crectness and linearity conformed to WS/T 406-2012 requirements.Conclusion The Beckman Coulter LH750 blood cell analyzer has good performance,its detection results are accurate and reliable,which can meet the requirements of clinical testing.
引文
[1]中国合格评定国家认可委员会.医学实验室质量和能力认可准则在临床血液学检验领域的应用说明:CNAS-CL02-A001[S].北京:中国合格评定国家认可委员会,2018.
[2]中华人民共和国卫生部.临床血液学检验常规项目分析质量要求:WS/T 406-2012[S].北京:中华人民共和国卫生部,2012.
[3]中华人民共和国卫生部.血细胞分析的校准指南:WS/T347-2011[S].北京:中华人民共和国卫生部,2011.
[4]NCCLS.Evaluation of the linearity of quantitaive measuremengt procedures:EP6-A2[S].Wayne,PA,USA:NCCLS,2003.
[5]张时民,王庚.血细胞分析自动化与显微镜复检应关注什么[J].国际检验医学杂志,2016,37(4):433-435.
[6]杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008.
[7]中国合格评定国家认可委员会.医学实验室质量和能力认可准则:ISO15189:2012[S].北京:中国合格评定国家认可委员会,2015.
[8]禚金花,刘振军.MEK-8222K血细胞分析仪性能验证及评价[J].国际检验医学杂志,2016,37(3):397-399.
[9]袁莉,陈葳,王刚.全自动血凝仪SYSMEX CS5100性能验证[J].检验医学与临床,2016,13(4):517-519.
[10]中华人民共和国卫生部.临床化学设备线性评价指南:WS/T 408-2012[S].北京:中华人民共和国卫生部,2013.