C-878定碳仪在铀铌合金中碳含量测定的应用
详细信息    查看官网全文
摘要
本文应用C-878红外定碳仪,对铀铌合金中碳进行了测定,考察了碳标准和铀铌样品在不同温度下对碳释放的影响。对铀铌样品的测量结果与LECO CS-444红外仪测量的结果做了比对实验,对于铀合金含碳为50μg/g,其相对标准不确定度<10%,加入标准的回收率为92%~106%;含碳为180μg/g,其相对标准不确定度<5%,加入标准的回收率为90%~105%。方法的测量下限为9μg/g。测定标准样品在1.25μg~12.5μg范围内,线性相关系数r为0.99986。
Carbon in U-Nb alloy was determined by the infrared carbon analyzer C-878.The effect of temperature on the releasing rate of carbon both in the standard and U-Nb alloy samples ware performed. The RSD is less than 10%, the recovery is between 92% and 106% respectively when carbon content in U-Nb alloy is about 50μg/g. The RSD is less than 5%, the recovery is between 90% and 105% respectively when carbon content in U-Nb alloy is about 180μg/g. The determination lower limit is 9μg/g. When the quantity of the standard samples ranges from 1.25μg~12.5μg,the linearity correlation coefficient is 0.99986.The results determined by are accord well with that by Leco CS-444.
引文
[1]万秀芝.使用红外分析仪测定超低碳[J].冶金分析,1995,15(3):54-55.
    [2]汤浩波.分析条件对碳硫分析仪线性的影响[C].LECO中国用户学术交流会,2001,47-49.
    [3]周昌震.红外气体分析仪的刻度曲线和谱带吸收.[J]分析仪器,1983,4:73-78.
    [4]邹乐西,吴伦强,向方寿,王林.铀铌合金中微量碳的测定[J].分析实验室,2001,20(3):79-81.
    [5]赵藻藩,周性尧,张悟铭,赵文宽.仪器分析[M].高等教育出版社,1990.