数字化地震前兆观测仪器维修
详细信息 本馆镜像全文    |  推荐本文 | | 获取馆网全文
摘要
主要介绍FHD-1型质子核旋磁力仪无补偿电流输出;RS-232C总线隔离驱动器(光隔副侧)引起的数据通信时好时坏;水位、地热数据不能调收;气氡、氦仪器无故障死机等故障的检修方法,对有关的电路及故障原因进行分析,并以实例对这些故障的检修进行了总结,提出了检修方法、步骤及改进措施。
In this article, the method for examining and repairing the FHD-1 proton nuclear spin magnetometer that no compensation electric current output, RS-232C bus isolated driver that makes data communication been on and off, water level, temperature data not been received, gas radon, helium instruments shut down without reason is introduced. The analysis on relative electric circuit and failures is also given. Examples are used for summarize the repair method, procedure and improving measures.
引文
李松阳等.2005.数字化地震仪系统的标定[J].地震地磁观测与研究,26(6)