A biomechanical study on suture anchor insertion angle: Which is better, 90° or 45°?
详细信息    查看全文
文摘
The purpose of this study was to compare the pullout strength of anchors inserted at 90° and 45° to the bone surface using synthetic bones and porcine humeri.Subjects and methodsPullout tests were performed by universal testing machine. Synthetic cancellous bones of 0.08, 0.16, and 0.24 g/cm3 (defining as low, medium and high density, respectively) with 2-mm-thick cortical bone model attached on one side and the greater tuberosity of porcine humeri (average bone density, 270 mg/cm3) were chosen for pullout tests. Metallic anchors were inserted at 90° or 45° to the surface and pulled at 90° or 45° from the surface. The maximum load to failure for each condition was recorded. Differences in pullout failure loads between insertion angle, pulling angle, and bone density were analyzed.ResultsWhen the sutures were pulled at 90° in low, medium, high density bones, and porcine humeri, 90°-inserted-anchors showed higher pullout strength than the 45°-inserted-anchors (534.6 ± 28.9 N vs. 488.1 ± 25.3 N (p < 0.05), 636.8 ± 25.3 N vs. 517.5 ± 27.4 N (p < 0.01), 735.6 ± 45.1 N vs. 557.0 ± 42.5 N (p < 0.01), and 285.6 ± 47.2 N vs. 181.4 ± 31.3 N (p < 0.01), respectively). When the sutures were pulled at 45° in low, medium density bones and porcine humeri, 90°-inserted-anchors showed higher pullout strength than the 45°-inserted-anchors (651.1 ± 38.3 N vs. 529.4 ± 37.6 N (p < 0.01), 711.4 ± 25.3 N vs. 599.2 ± 29.8 N (p < 0.01), and 265.3 ± 49.0 N vs. 181.5 ± 29.4 N (p < 0.01), respectively).ConclusionPullout strength of the anchors inserted at 90° to the bone surface was greater than the anchors inserted at 45° regardless of the bone density.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700