Preliminary investigation of scintillator materials properties: SrI2:Eu, CeBr3 and GYGAG:Ce for gamma rays up to 9 MeV
详细信息    查看全文
文摘
In this work we measured the performance of a 2ʺ×2ʺ cylindrical tapered crystal of SrI2:Eu, a 2ʺ×3ʺ cylindrical sample of CeBr3 and a 2ʺ×0.3ʺ cylindrical sample of GYGAG:Ce. These scintillators are prototypes in volume or material and were provided by the Lawrence Livermore National Laboratory and by the Institut de Physique Nucléaire d׳Orsay. The gamma-ray energy resolution was measured in the energy range of 0.1–9 MeV using different sources. Each scintillator was scanned along x, y and z axes, using a 400 MBq collimated 137Cs source. Owing to the GYGAG:Ce thickness, it was not possible to obtain the value of the energy resolution at 9 MeV and to scan the crystal along the z axis. The 662 keV full energy peak position and its FWHM were measured relative to the full energy peaks positions produced by a non-collimated 88Y source. The signals of the detectors were additionally digitized and compared, up to 9 MeV, using a 12 bit LeCroy 600 MHz oscilloscope.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700