The biomechanical effect of pedicle screw hubbing on pullout resistance in the thoracic spine
详细信息查看全文 | 推荐本文 |
摘要

Background context

The biomechanical fixation strength afforded by pedicle screws has been strongly correlated with bone mineral density. It has been postulated that 鈥渉ubbing鈥?the head of the pedicle screw against the dorsal laminar cortex provides a load-sharing effect, thereby limiting cephalocaudad toggling and improving the pullout resistance of the pedicle screw.

Purpose

To evaluate the pullout strength (POS) of monoaxial hubbed pedicle screws versus standard fixation in the thoracic spine.

Study design

Biomechanical investigation.

Methods

Twenty-two human cadaveric thoracic vertebrae were acquired and dual-energy X-ray absorptiometry scanned. Osteoporotic (n=16) and normal (n=6) specimens were instrumented with a 5.0脳35-mm pedicle screw on one side in a standard fashion. In the contralateral pedicle, 5.0脳30-mm screw was inserted with hubbing of the screw into the dorsal lamina. A difference in screw length was used to achieve equivalent depth of insertion. After 2,000 cycles of cephalocaudad toggling, screws were pulled out with the tensile force oriented to the midline of the spine and peak POS measured in newtons (N). Four additional specimens were subjected to microcomputed tomography (micro-CT) analysis to evaluate internal pedicle architecture after screw insertion.

Results

Hubbed screws resulted in significantly lower POS (290.5卤142.4 N) compared with standard pedicle screws (511.5卤242.8 N; p=.00). This finding was evident in both normal and osteoporotic vertebrae based on independent subgroup post hoc analyses (p<.05). As a result of hubbing, half of the specimens fractured through the lamina or superior articular facet (SAF). No fractures occurred on the control side. There was no difference in mean POS for hubbed screws with and without fracture; however, further micro-CT analysis revealed the presence of internal fracture propagation for those specimens that did not have any external signs of failure.

Conclusions

Hubbing pedicle screws results in significantly decreased POS compared with conventional pedicle screws. Hubbing predisposes toward iatrogenic fracture of the dorsal lamina, transverse process, or SAF during insertion.