Clinical evaluation of an end-tidal target-controlled infusion closed-loop system for isoflurane administration in horses undergoing surgical procedures
详细信息查看全文 | 推荐本文 |
摘要
A new volatile anaesthetic agent delivery system was tested in 15 horses undergoing scheduled surgical procedures. The delivery system consisted of a laptop computer (with dedicated software), a computer-controlled syringe driver (loaded with liquid isoflurane) connected to the inspiratory arm of a large-animal circle breathing system and a respiratory gas monitor, providing isoflurane end-tidal concentrations (ETmeasured) every 20 s to the computer. Following induction and connection to the breathing system, mechanical ventilation was started. The bodyweight (BW), fresh gas flow, breathing system and ventilator volume, and end-tidal isoflurane target (ETtarget) were entered into the computer. Using Lowe鈥檚 equation, the software calculated the prime dose to be delivered by the syringe driver over 2 min. After this, the system delivered each minute the amount of isoflurane as determined by the following equation: Isoflurane (mL) = {2 脳 B/G 脳 (200 脳 BW0.75) 脳 (ETtarget 鈭?#xA0;ETmeasured) + (fresh gas flow 鈭?#xA0;(BW0.75 脳 0.07)) 脳 (ETmeasured)}/206. A fresh gas flow of 4 L oxygen min鈭? was administered until the inspired fraction of oxygen reached 0.7, and was then decreased. A target of 1.5%end-tidal isoflurane was initially used and subsequently adjusted to the clinical requirements. The system performance was evaluated using the median prediction error (MDPE) and the median absolute performance error (MDAPE), which were 鈭?.6%and 5.29%, respectively. It was concluded that this system was useful to achieve end-tidal target-controlled infusion of isoflurane during equine anaesthesia.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700