用户名: 密码: 验证码:
Determination of film thickness and surface profile using reflectometry and spectrally resolved phase shifting interferometry
详细信息
下载全文
推荐本文 |
摘要
Surface profiling and film thickness measurement play an important role for inspection in semi conductor industry. White light source had been used as scanning white light interferometry and spectrally resolved white light interferometry for determining surface and film thickness profile. These techniques however failed for thinner film. Recently, reflectometry and spectrally resolved white light interferometry was combined for the same. This technique used Fourier Transform for the calculation of phase in spectral domain with the use of Linnik interferometer. In this method a large amount of carrier offset (carrier fringes) is required to be effective. This carrier fringes in spectrally resolved white light interferometry was achieved by increasing the optical path difference between the test and the reference surface. But, Linnik interferometer cause defocusing problem to create these carrier fringes. We propose in this paper to combine reflectometry and spectrally resolved phase shifting interferometry for measurement of surface and film thickness profile with the use of Michelson objective. Michelson objective will be convenient to implement as compared to the Linnik type and the use of phase shifting interferometry does not necessarily need large number of fringes in the spectral domain.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700