用户名: 密码: 验证码:
基于容积脉搏波振幅梯度的血压检测法的研究
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:A method of blood pressure measurement based on the amplitude gradient of volume pulse wave
  • 作者:宋义林 ; 张彤 ; 高树枚
  • 英文作者:SONG Yilin;ZHANG Tong;GAO Shumei;School of Mechanical & Electrical Engineering,Heilongjiang University;
  • 关键词:血压 ; 容积脉搏波振幅梯度 ; 血管壁力学特性 ; 容积脉搏波 ; 检测系统
  • 英文关键词:Blood pressure;;Amplitude gradient of volume pulse wave;;Mechanical characteristics of vessel wall;;Volume pulse wave;;Measurement system
  • 中文刊名:SDSG
  • 英文刊名:Journal of Biomedical Engineering Research
  • 机构:黑龙江大学机电工程学院;
  • 出版日期:2018-09-25
  • 出版单位:生物医学工程研究
  • 年:2018
  • 期:v.37
  • 基金:黑龙江省自然科学基金资助项目(C2015025);; 黑龙江省教育厅科学技术研究重点项目(HDJCCX-2016209)
  • 语种:中文;
  • 页:SDSG201803008
  • 页数:6
  • CN:03
  • ISSN:37-1413/R
  • 分类号:43-47+58
摘要
提出了基于容积脉搏波振幅梯度的血压检测方法,并开发了相应的检测系统,进行了原理测试和对比实验。首先,对三名实验者进行了验证性实验,确认了系统的稳定性;其次,以水银式血压检测法作为对比方法,进行了基于容积脉搏波振幅梯度的血压检测法的对比实验,实验结果表明两种血压检测法具有良好的直线相关性;最后,用Bland-Altman图进行分析,结果表明收缩压和舒张压均有94%以上的实验结果分布在95%置信区间的范围之内,说明两种测量方法具有良好的一致性。基于容积脉搏波振幅梯度的血压检测法与水银式血压检测法具有大致相同的检测精度,本研究提出的新的血压检测法对间歇式血压检测是有效的。
        A new blood pressure measurement method based on the amplitude gradient of volume pulse wave was proposed,a corresponding detection system was developed,and the principle test and the comparison experiment for the system were carried out. Firstly,a confirmatory test was conducted by three participants involved,and the stability of the system was confirmed. Secondly,a contrast experiment of the blood pressure measurement method proposed and the mercury sphygmomanometer blood pressure measurement method was carried out. The experimental results showed that two kinds of blood pressure detection method had good linear correlation. Finally,the Bland-Altman method was used to analyze the consistency of two kinds of detection methods. The experimental results showed that 94% of the detection data of the systolic blood pressure and diastolic blood pressure was located in the 95% confidence interval range,showing two kinds of measurement method with good consistency. The blood pressure detection method based on the amplitude gradient of volume pulse wave has approximately the same detection accuracy as the mercury sphygmomanometer blood pressure measurement method,and the new method presented in this paper is effective for the detection of blood pressure.
引文
[1]陈伟伟,高润霖,刘力生,等.中国心血管病报告2014概要[J].中国循环杂志,2015,30(7):617-622.
    [2]Pan F,Zheng D,Murray A.Effect of stethoscope position on auscultatory blood pressure measurement[C]//Computing in Cardiology Conference,Zaragoza,2013:895-898.
    [3]Liu,Jian Kun,Mukkamala,et al,Error mechanisms of the oscillometric fixed-ratio blood pressure measurement method[J].Annals of Biomedical Engineering,2013,41(3):587-597.
    [4]Forouzanfar M,Dajani H R,Groza V Z,et al.Oscillometric blood pressure estimation:past,present,and future[J].IEEE Reviews In Biomedical Engineering,2015,8:44-63.
    [5]Lee S,Chang J H.Oscillometric blood pressure estimation based on deep learning[J].IEEE Transactions On Industrial Informatics,2016,pp(99):1-1.
    [6]Lee S,Chang J H,Sang W N,et al.Oscillometric blood pressure estimation based on maximum amplitude algorithm employing gaussian mixture regression[J].IEEE Transactions on Instrumentation and Measurement,2013,62(12):3387-3389.
    [7]Ando S I,What does a perfect blood pressure meter look like from a clinician point of view?[J].IEEE Instrumentation&Measurement Magazine,2014,17(3):15-20.
    [8]白丽红,王成,文苗,等。基于脉搏波传导时间的连续血压监测系统[J].生物医学工程研究,2014,33(4):221-225.
    [9]包科,顾亚雄,高俊,等。基于脉搏波传导时间的连续血压测量系统[J].生物医学工程研究,2016,35(1):41-45.
    [10]宋义林,高树枚.基于局部压迫法的指动脉血压连续检测用新型检测装置开发[J].黑龙江大学工程学报,2010,1(4):104-109.
    [11]钱德兰,里特杰斯,尤甘纳坦.生物流体力学:人体循环系统[M].机械工业出版社,2015,107-116.
    [12]Bland J M,Altman D G,Statistical methods for assessing agreement between two methods of clinical measurement[J].Lancet,1986,327(8476):307-310.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700