摘要
采用正己烷提取水中的邻苯二甲酸二丁酯,并建立水中痕量邻苯二甲酸二丁酯的测定方法。NIST标准谱库检索定性,SIM模式下特定离子外标法定量。结果表明:邻苯二甲酸二丁酯浓度在10ug/L-500ug/L范围内呈现良好的线性关系,相关系数等于0.9996;样品加标回收率在86.7%-99.5%之间,方法的检出限为0.1ug/L。能满足水中痕量邻苯二甲酸二丁酯测定的要求。
To determine trace amount of DBP in surface water, a method was developed using gas chromatography/mass spectrometry with n-hexane as extraction solvent. NIST standard spectral library search for qualitative, SIM mode specific ion external for quantitative. good relativity was obtained in the range of 10-500 ug/L, The related coefficient of calibration curve was 0.9996; The recoveries are between 86.7%-99.5%,The determination limit is less than 0.1ug/L,The method is suitable for the determination of DBP in water.
引文
[1]孙宝国,孙金沅.邻苯二甲酸酯类化合物的天然存在及安全性问题的探讨[J].中国食品学报,2011,11(8):1-8.
    [2]GB 3838-2002.地表水环境质量标准[S],2002.
    [3]韩敏,樊晓涛,郭艳利,等.高效液相色谱法测定简易塑料手套中邻苯二甲酸二丁酯的研究[J].应用化工,2015,44(11):2018-2021.
    [4]王晓红,曾芃斐,余志铨,等.超高效液相色谱-质谱法测定饮用水源水中邻苯二甲酸酯类化合物[J].绿色科技,2016(10):77-79.
    [5]王建娥,苏凤仙,杨玲,等.气相色谱法测定工业废水中邻苯二甲酸二丁酯含量[J].合成技术及应用,2016,31(3):48-52.