用户名: 密码: 验证码:
Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow LCMSn
详细信息    查看全文
  • 作者:Dhanashri BagalEddie Kast ; Ping Cao
  • 刊名:Analytical Chemistry
  • 出版年:2017
  • 出版时间:January 3, 2017
  • 年:2017
  • 卷:89
  • 期:1
  • 页码:720-727
  • 全文大小:501K
  • ISSN:1520-6882
文摘
Monoclonal antibodies (mAbs) are large heterogeneous molecules that represent a growing class of therapeutics. De novo sequencing of mAbs becomes necessary when the original cell line or the cDNA is unavailable. An important feature in sequencing of mAbs is the discrimination of isobaric residues (Xle): leucine (Leu) and isoleucine (Ile). An incorrect identification of the Xle site, especially in the complementarity determining regions (CDRs), can result in the production of an antibody with severely compromised efficacy. Multistage fragmentation (MSn) in the mass spectrometer can provide sufficient evidence for Ile/Leu discrimination. However, most existing methods utilize direct infusion of purified peptides, demanding peptide enrichment which can be labor-intensive and requires large amount of material. Here we introduce an online nano-LCMSn method, which depending on the nature of the peptide, exploits either generation of a signature 69 Da ion from Ile or formation of unique w-ions employing MS3 (ETD-HCD) for rapid Ile/Leu distinction. This reliable and sensitive method utilizes the Orbitrap Fusion tribid mass spectrometer to rapidly assign multiple Xle residues in the CDRs of mAbs.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700