Comparison of Sebia Capillarys Flex capillary electrophoresis with the BioRad Variant II high pressure liquid chromatography in the evaluation of hemoglobinopathies
详细信息查看全文 | 推荐本文 |
摘要

Background

There are multiple biochemical screening techniques for assessing hemoglobinopathies. Here we compare a new instrument, the Sebia Capillarys Flex (capillary zone electrophoresis (CE)), with the BioRad Variant II (high pressure liquid chromatography (HPLC) in the evaluation of hemoglobinopathies.

Methods

This was a retrospective study using 174 whole blood samples encompassing the 5 most common (Hb A, A2鈥? S, C, and E) and 10 rare (Hb GPhilidelphia, D, H, Bart's, OArab, S/GPhilidelphia, Hasharoon, QIndia, NBaltimore, and Malmo) hemoglobin variants. An additional 126 samples were used to establish a CE reference interval for Hb A2.

Results

Hb A measurements agreed well between the 2 methods (bias = 鈭?#xA0;0.06;r = 0.999). The agreement of Hb F was also very good (bias = 鈭?#xA0;0.17; r = 0.994). When samples with the highest Hb F concentration were excluded, agreement was less precise (bias = 鈭?#xA0;0.44; r = 0.811). When no variant was present, the Hb A2 concentrations showed excellent agreement (bias = 0.00; r = 0.994). Positive bias for Hb A2 is seen when Hb C is present using CE. The Hb A2 reference interval using CE was < 3.2%.

Conclusion

The Capillarys Flex is capable of identifying and quantifying hemoglobin species, consistent with existing HPLC methods.