用户名: 密码: 验证码:
GEM-TEPC调研分析及原理研究
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Survey on GEM-TEPC and Study on its Principle
  • 作者:李春娟 ; 张伟华 ; 肖雪夫 ; 王志强 ; 刘毅娜
  • 英文作者:LI Chun-juan;ZHANG Wei-hua;XIAO Xue-fu;WANG Zhi-qiang;LIU Yi-na;National Key Laboratory for Metrology and Calibration Techniques,China Institute of Atomic Energy;
  • 关键词:气体电子倍增技术 ; 组织等效正比计数器 ; 微剂量学 ; 结构设计
  • 英文关键词:Gas Electron Multiplier(GEM);;Tissue-Equivalent Proportional Counter(TEPC);;Microdosimetry;;Structure design
  • 中文刊名:YHJJ
  • 英文刊名:Journal of Astronautic Metrology and Measurement
  • 机构:中国原子能科学研究院计量与校准技术重点实验室;
  • 出版日期:2019-02-15
  • 出版单位:宇航计测技术
  • 年:2019
  • 期:v.39;No.229
  • 语种:中文;
  • 页:YHJJ201901006
  • 页数:6
  • CN:01
  • ISSN:11-2052/V
  • 分类号:25-29+73
摘要
为对现有组织等效正比计数器进行改进,使其满足高注量率辐射场测量要求,本实验室拟研制一套基于气体电子倍增技术的宽量程TEPC。对GEM-TEPC工作原理进行研究,并对其国际发展现状进行调研,在传统的TEPC灵敏体积内引入GEM膜取代阳极丝是切实可行的,可降低微型TEPC探测器的加工难度,提高其性能。通过调研总结GEM-TEPC设计关键技术,为本实验室GEM-TEPC的研制奠定了一定基础。
        To optimize the design of tissue-equivalent proportional counter( TEPC) and make it satisfy the measurement of radiation field with high fluence rate,principles of GEM-TEPC was studied,and its development status were surveyed on. The results show that it is available to use GEM instead of anode wire to decrease the fabrication difficulty of mini-TEPC and improve characteristics of TEPC. Finally,the key technology of GEM-TEPC were studied,which lays foundation of the development of GEM-TEPC in our lab.
引文
[1]Moro D,ColauttiP,LolloM,et al.BNCT dosimetry performed with a mini twin tissue-equivalent proportional counters(TEPC)[J].Applied Radiation and Isotopes,2009,67:171-174.
    [2]Aslam,Waker A J.Performance of a high sensitivity multi-element tissue equivalent proportional counter for radiation protection neutron monitoring measurements[J].Health Phys,2010,98(5):692-703.
    [3]Waker A J,Aslam,?Lori J,et al.Design of a multielement tepc for neutron monitoring[J].Radiation Protection Dosimetry,2011,143(2-4):463-466.
    [4]SAULI F.GEM:a new concept for electron amplification in gas detectors.Nuclear Instrument and Method A,1997,386:531-534.
    [5]Rossi H H,Zaider M.Microdosimetry and its applications[J].Physics today,1997,50(12):70-71.
    [6]田志恒.辐射剂量学[M].北京:原子能出版社,1992:126-127.
    [7]M,Bos A J J,van Eijk C W E.Gas electrical multiplier(GEM)operation with tissue equivalent gases at various pressures[J].Nuclear Instrument and Method A,2003,506:160-165.
    [8]M.Farahmand.A novel Tissue-Equivalent Proportional Counter Based on a Gas Electrical Multiplier[D].Delft:Delft University of Technology,2004.
    [9]Chris Wang.Development and Test of a GEM-Based TEPC System for In-Phantom Dose Measurements[D].Atlanta:Georgia Institute of Technology,2007.
    [10]G.M.Orchard,K.Chin,W.V.Prestwic,et al.Development of a thick gas electron multiplier for microdosimetry[J].Nuclear Instrument and Method A,2011,638:122-126.
    [11]张伟华,王志强.组织等效正比计数器的测量原理和方法[J].剂量研究,2008(3):45-54.
    [12]张伟华,王志强.组织等效正比计数器测量系统的建立[J].原子能科学技术,2010,44(11):1830-1834.
    [13]张伟华,王志强.组织等效正比计数器结构优化设计[J].原子能科学技术,2014,48(5):930-933.
    [14]李春娟张伟华肖雪夫王志强.GEM-TEPC电场模拟及增益分析方法研究[J].宇航计测技术,2017,1:72-75

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700