电感耦合等离子体质谱仪双样品环进样系统的设计
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Design of Double Sampling Loops System of Inductively Coupled Plasma Mass Spectrometry
  • 作者:石雅静 ; 刘宏 ; 王彤 ; 徐进勇 ; 向梅 ; 任子位
  • 英文作者:SHI Ya-jing;LIU Hong;WANG Tong;XU Jin-yong;XIANG Mei;REN Zi-wei;College of Earth Sciences, Chengdu University of Technology;College of Materials and Chemistry & Chemical Engineering,Chengdu University of Technology;
  • 关键词:双样品环进样系统 ; 电感耦合等离子体质谱仪(ICP-MS) ; 六通多路阀 ; 测试效率
  • 英文关键词:double sampling loops system;;inductively coupled plasma mass spectrometer(ICP-MS);;switching valve;;detection efficiency
  • 中文刊名:ZPXB
  • 英文刊名:Journal of Chinese Mass Spectrometry Society
  • 机构:成都理工大学地球科学学院;成都理工大学材料与化学化工学院;
  • 出版日期:2019-04-17 09:24
  • 出版单位:质谱学报
  • 年:2019
  • 期:v.40
  • 基金:国家自然科学基金项目(41873014)资助
  • 语种:中文;
  • 页:ZPXB201903006
  • 页数:6
  • CN:03
  • ISSN:11-2979/TH
  • 分类号:52-57
摘要
电感耦合等离子体质谱仪(ICP-MS)常搭配溶液气动雾化器进样系统,以串联方式测样,其进样效率并未达到最佳。本文基于六通多路阀设计了一种双样品环进样系统,连接雾化器前,用以提高仪器的测样效率。实验测定了Sc、Ti、V、Cr、Mn、As、Co、Ni、Cu、Pb、Zn、Cd、Li、Sr、Ba、Co、U、Th共18种元素,其校准曲线的线性相关系数均大于0.999,精密度(n=5)均小于5%。实验还分别测定了1、6、12、18、32种元素,测样时间分别节约了45%、35%、27%、21%、16%。以Ce元素为例,先测试100.00μg/L样品后,立即测试1.00μg/L样品,未发现明显的记忆效应。该系统价格低廉、操作简单、体积小、易于实现自动化,且有与其他仪器联用的可能性,具有广阔的发展前景。
        Inductively coupled plasma mass spectrometry(ICP-MS) is usually coupled with solution pneumatic nebulizer sampling system, whose detection efficiency is low when measured in series mode. Based on the double-layer six-way multi-way valve, a double-sample ring sampling system was designed, which could improve the detection efficiency by changing the connection mode of the sampling system. This system connected with pneumatic nebulizer, one loop was loading sample, the other loop was detecting. The detection time was saved by two-way parallel mode. Determination of different numbers of elements needed different length loop. The correlation coefficients of calibration curves of Sc, Ti, V, Cr, Mn, As, Co, Ni, Cu, Pb, Zn, Cd, Li, Sr, Ba, Co, U, and Th are greater than 0.999, and the precisions(n=5) are less than 5%. 1, 6, 12, 18, and 32 elements were determinted, their detection time was saved 45%, 35%, 27%, 21%, 16%, respectively. The number of elements is fewer, the improvement of the efficiency is higher. Compared with standard method, the limits of detection and sensitivity of this system have no significant difference. The double sampling loops system has no significant memory effect. The two-layer six-way multi-way valve switched manually and owned automation prospect. This system is high efficient, fast and simple, easy to realize automation, and more possibilities to use with other instruments.
引文
[1] 郭冬发,李金英,李伯平,谢胜凯,谭靖,张彦辉,刘瑞萍.电感耦合等离子体质谱分析方法的重要进展(2005~2016年)[J].质谱学报,2017,38(5):599-610.GUO Dongfa,LI Jinying,LI Boping,XIE Shengkai,TAN Jing,ZHANG Yanhui,LIU Ruiping.Important progress in inductively coupled plasma mass spectrometry (2005-2016)[J].Journal of Chinese Mass Spectrometry Society,2017,38(5):599-610(in Chinese).
    [2] 李金英,郭冬发,姚继军,曹淑琴.电感耦合等离子体质谱(ICP-MS)新进展[J].质谱学报,2002,23(3):164-169.LI Jinying,GUO Dongfa,YAO Jijun,CAO Shuqin.New progress in inductively coupled plasma mass spectrometry (ICP-MS)[J].Journal of Chinese Mass Spectrometry Society,2002,23(3):164-169(in Chinese).
    [3] 丁兰,金海燕,刘淼,梁枫,郇延富,曹彦波,张寒琦,金钦汉.用低功率微波热雾化器与十字交叉气动雾化器作为ICP-AES进样装置的比较[J].高等学校化学学报,2000,21(11):1 661-1 663.DING Lan,JIN Haiyan,LIU Miao,LIANG Feng,XUN Yanfu,CAO Yanbo,ZHANG Hanqi,JIN Qinhan.Comparison of low power microwave thermal atomizer and cross-pneumatic atomizer as ICP-AES sampling device[J].Chemical Journal of Chinese Universities,2000,21(11):1 661-1 663(in Chinese).
    [4] 白岗,孔祥丽,霍阳.一种雾化去溶捕集样品引入系统及方法:CN,CN 103512944A[P].2014.
    [5] 李冰,杨红霞.电感耦合等离子体质谱仪原理和应用[M].北京:地质出版社,2005.
    [6] Tsuyoshi Morikawa,Tsutomu Okoba.Autosampler:USA.US09638674B2[P].2017-05-02.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700