Evaluation of Uncertainty for Determining Iodide in Groundwater by Ion Chromatography with Integrated Amperometric Detection
详细信息   
摘要

The measurement uncertainty of determining the content of iodide in groundwater samples by ion chromatography with integrated amperometric detection was evaluated in this paper. The calculating model of double error fitting in the course of standard curve fitting was used, and the mathematical model of uncertainty was also established to obtain some factors affecting those results.