Spectroscopic Analysis of Trace Ag in Geochemical Sample
详细信息   
摘要
Taking Cd chloride and ammonia chloride as the reaction agent, background as inner standard, the authors analyzed the trace Ag of geochemical sample with WSP-1 plane grating spectrometer.  The linear range of method is 0.025 μg/ml.  The relative error of 6 teats is 1.68%3.39%.  In the test of standard sample, the testing value is basically consistent with the recommendation value, which can meet the basic requirement in testing trace Ag in geochemical sample.