Influence of mobile phase, source parameters and source type on electrospray ionization efficiency in negative ion mode
详细信息    查看全文
  • 作者:Anneli Kruve
  • 刊名:Journal of Mass Spectrometry
  • 出版年:2016
  • 出版时间:August 2016
  • 年:2016
  • 卷:51
  • 期:8
  • 页码:596-601
  • 全文大小:238K
  • ISSN:1096-9888c
文摘
Electrospray ionization (ESI) efficiency is known to be affected by the properties of the analytes, source design and source parameters. In this study, the ionization efficiency of 17 acidic compounds at various conditions in ESI negative ion mode was evaluated. Namely, the influence of organic solvent content in the mobile phase, ionization source parameters, ionization source geometry and functionality (conventional ESI, ESI with thermal focussing and with additional internal nebulizer gas) was studied. It was observed that the ionization efficiency in thermal focussing ESI is only marginally affected by the organic solvent composition, while for conventional ESI and ESI with internal nebulizer gas, the ionization efficiency increases significantly with increasing organic modifier content. For all ionization sources and mobile phase compositions, the ionization efficiency values between different setups showed good correlation. Copyright