Cardiac Index Declines During Long-Term Left Ventricular Device Support
详细信息    查看全文
文摘
To investigate longitudinal trends in valvular and ventricular function with long-term left ventricular assist device (LVAD) therapy, we analyzed hemodynamic and echocardiographic data of patients with at least 2 years of continuous LVAD support. All 130 patients who underwent HeartMate II implantation at our institution between 2005 and 2012 were reviewed. Twenty patients had hemodynamic and echocardiographic evaluations in both the early (0–6 months) and late (2–3 years) postoperative period. Patients on inotropic therapy or temporary mechanical support were excluded. The average times of early and late hemodynamic evaluations were 59 ± 41 days and 889 ± 160 days, respectively. Cardiac index (CI) declined by an average of 0.4 L/min/m2 (P = 0.04) with concomitant increase in pulmonary capillary wedge pressure (PCWP; P = 0.02). The right atrial pressure to PCWP (RAP:PCWP) ratio decreased during LVAD support suggesting improvement in right ventricular function. While there was an increase in degree of aortic insufficiency (AI) at the late follow-up period (P = 0.008), dichotomization by median decline in CI (−0.4 L/min/m2) indicated no difference in prevalence of AI among the groups. CI declined in patients with HeartMate II after 2 years of continuous support. An increase in preload and afterload was observed in those with the greatest decline in CI.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700