Characterisation of horseradish peroxidase immobilisation on an electrochemical biosensor by colorimetric and amperometric techniques
详细信息    查看全文
文摘
This study presents the use of complementary colorimetric and amperometric techniques to measure the quantity of protein or enzyme immobilised onto a carbon paste electrode modified with a layer of electrodeposited polyaniline. By applying a solution of bovine serum albumin at 0.75 mg/ml, efficient blocking of the electrode from electroactive species in the bulk solution could be achieved. When the horseradish peroxidase was immobilised on the electrode, optimal amperometric responses from hydrogen peroxide reduction were achieved at approximately the same concentration. The mass of enzyme immobilised at this solution concentration was determined by a colorimetric enzyme assay to be equivalent to the formation of a protein monolayer. Under these conditions, amperometric responses from the immobilised layer are maximised and non-specific bulk solution interactions are minimised. At higher immobilised protein concentrations, diminished amperometric responses may be due to inhibited diffusion of hydrogen peroxide to enzyme which is in electronic communication with the electrode surface, or impeded electron transfer.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700