A flow-batch luminometer
详细信息    查看全文
文摘
The traditional flow chemiluminescence (CL) systems typically use flow cells in spiral shape mounted in front of a photomultiplier tube (PMT). The sample is transported by using carrier fluids, undergoes dispersion, and mixes with reagents and CL reaction occurs before reaching the flow cell. As a consequence, the maximum of the chemiluminescence intensity is not exploited and the sensitivity is significantly diminished. In this work, these drawbacks were overcome by using a flow-batch approach and a large area silicon photodiode instead of PMT in order to build a simple and automatic luminometer for CL measurements. The feasibility of the proposed flow-batch luminometer was demonstrated in the determination of vitamin B12 in injection ampoules, yielding limits of detection and quantification (0.11 and 0.36 ¦Ìg L? 1, respectively), recovery rates (between 97.8 and 102.1 % ), relative standard deviations (RSD < 2.2 % , n = 3) and sample throughput (about 72 h? 1). Thus, it was possible to both project and build a simple, flexible, versatile and automatic luminometer, while keeping the excellent characteristics of the previous flow-batch analyzer such as: low reagent and sample consumption and minimal waste generation.