Evaluation of the user seal check on gross leakage detection of 3 different designs of N95 filtering facepiece respirators
详细信息    查看全文
文摘
The use of N95 respirators prevents spread of respiratory infectious agents, but leakage hampers its protection. Manufacturers recommend a user seal check to identify on-site gross leakage. However, no empirical evidence is provided. Therefore, this study aims to examine validity of a user seal check on gross leakage detection in commonly used types of N95 respirators.

Methods

A convenience sample of 638 nursing students was recruited. On the wearing of 3 different designs of N95 respirators, namely 3M-1860s, 3M-1862, and Kimberly-Clark 46827, the standardized user seal check procedure was carried out to identify gross leakage. Repeated testing of leakage was followed by the use of a quantitative fit testing (QNFT) device in performing normal breathing and deep breathing exercises. Sensitivity, specificity, predictive values, and likelihood ratios were calculated accordingly.

Results

As indicated by QNFT, prevalence of actual gross leakage was 31.0%-39.2% with the 3M respirators and 65.4%-65.8% with the Kimberly-Clark respirator. Sensitivity and specificity of the user seal check for identifying actual gross leakage were approximately 27.7% and 75.5% for 3M-1860s, 22.1% and 80.5% for 3M-1862, and 26.9% and 80.2% for Kimberly-Clark 46827, respectively. Likelihood ratios were close to 1 (range, 0.89-1.51) for all types of respirators.

Conclusions

The results did not support user seal checks in detecting any actual gross leakage in the donning of N95 respirators. However, such a check might alert health care workers that donning a tight-fitting respirator should be performed carefully.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700