A new series of uranium isotope reference materials for investigating the linearity of secondary electron multipliers in isotope mass spectrometry
详细信息    查看全文
文摘
A new series of gravimetrically prepared uranium isotope reference materials, the so-called IRMM-074 series, with the n(235U)/n(238U) isotope ratio held constant at unity and the n(233U)/n(238U) isotope ratios varying from 1.0 to 10−6 has been prepared and certified. This series is suited for calibration of secondary electron multipliers used widely in isotope mass spectrometry, in particular for techniques such as thermal ionization mass spectrometry (TIMS), inductively coupled plasma mass spectrometry (ICPMS), accelerator mass spectrometry (AMS) and resonance ionization mass spectrometry (RIMS). The new IRMM-074 was prepared as a replacement for the already exhausted IRMM-072 predecessor series.

Uranium materials with high isotopic enrichments of 233U, 235U and 238U were purified using identical methods involving separation on anion and cation column followed by a precipitation as peroxide. The oxides were calcined to convert them to U3O8 simultaneously, in an oven installed in a glove-box that provided a controlled low-humidity environment. The oxides of 235U and 238U were weighed and mixed with a mole ratio n(235U)/n(238U) = 1.0 and then dissolved. The 233U oxide was dissolved to form a separate solution with the same concentration and from this primary solution three dilutions were made by weighing. A weighed amount of the n(235U)/n(238U) solution and weighed amounts of the 233U solutions were mixed in various proportions in order to achieve n(233U)/n(238U) isotope ratios varying from 1.0 to 10−6. The methods for the preparation, the mixing and the mixing calculations are described. The expanded uncertainties (coverage factor k = 2) of the certified isotope ratios for the IRMM-074 series are 0.015 % for the n(235U)/n(238U) ratio and 0.025 % for the n(233U)/n(238U) ratios, which constitutes an improvement compared to those of the predecessor IRMM-072 series.

In addition, recent observations regarding the linearity response of secondary electron multipliers (SEMs) and suitable reference materials for investigating detector linearity are reviewed. Two measurement procedures for applying the IRMM-072 and IRMM-073 (diluted from the remaining fraction of IRMM-072) series as well as the new IRMM-074 series for assessing SEM linearity are suggested. The procedures are tailor-made for the specific instrumental characteristics of thermal ionization mass spectrometers (TIMS) and multiple-collector inductively coupled plasma mass spectrometers (MC-ICPMS) but can be adapted also for further types of isotope ratio mass spectrometers.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700