Evaluaci¨®n del sistema automatizado UF-1000i? en el diagn¨®stico de infecci¨®n urinaria
详细信息    查看全文
文摘

Objective

To evaluate the Sysmex UF-1000i? system as a urine screening method for the diagnosis of urinary tract infection, and epithelial cells as a predictive value of contamination in woman of childbearing age.

Methods

A total of 1730 urine samples were processed using a urine culture as a reference.

Results

With 50 bacteria/¦Ìl as a cut-off point, the results were: sensitivity 91.3 % , specificity 73.1 % , negative predictive value 96.2 % . For a specificity of 90 % for epithelial cells, the results were: sensitivity 31.0 % , positive predictive value 67.0 % , negative predictive value 66.0 % .

Conclusion

The evaluated system is fast and effective. Epithelial cells could be used to predict contamination.