The development of devices with internal gain, avalanche photodiodes (APD) first and then Geiger-mode avalanche photodiodes, named single photon avalanche diode (SPAD), leads to a substantial improvement in sensitivity and allowed single photon detection. Later on, thousands of SPADs have been assembled in arrays of few millimeters squared (named SiPM, silicon photo-multiplier) with single photon resolution.
The high internal gain of SiPMs, together with other features peculiar of the silicon technology like compactness, speed and compatibility with magnetic fields, promoted SiPMs as the principal photo-detector competitor of photomultipliers in many applications from radiation detection to medical imaging.
This paper provides a review of the properties of analog solid-state photo-detectors. Particular emphasis is given to latest advances on Positron Emission Tomography instrumentation boosted by the adoption of the silicon photo-detectors as an alternative to photomultiplier tubes (PMTs). Special attention is dedicated to the SiPMs, which are playing a key role in the development of innovative scanners.