The Macular Automated Photostress Test
详细信息    查看全文
文摘

Purpose

To introduce a standardized macular photostress test using an automated perimeter as a method to quantify macular disease severity and as a tool to distinguish optic neuropathy from macular pathology.

Design

Prospective interventional pilot study.

Methods

Twenty-five bilaterally pseudophakic subjects aged range, 65 to 84: 15 patients with varying severity of non-neovascular age-related macular degeneration (AMD), five patients with no ocular disease, and five patients with moderate primary open-angle glaucoma (POAG). Previously reported normative values served as controls for this study. Patients underwent foveal threshold testing using the Humphrey Visual Field Perimeter Model 750 (Carl Zeiss Meditec, Inc, Dublin, California, USA). Baseline measurements were compared to threshold sensitivity after photostress at one minute and then two-minute intervals until sensitivity returned to baseline. Main outcome measures were baseline foveal threshold sensitivity, foveal threshold depression, and recovery following photostress.

Results

Automated macular photostress testing in macular disease (AMD) causes a decrease (P < .001) in baseline foveal sensitivity and a delay (P < .001) in recovery time to baseline sensitivity. Optic nerve pathology (POAG) does not affect (P = .343) the foveal response curve.

Conclusions

The macular automated photostress (MAP) test is an inexpensive, noninvasive, and readily accessible adjunct for evaluating patients with macular disease. This standardized protocol is useful in objectively defining disease severity, may be used to follow response to treatment, and could aid in distinguishing optic neuropathy from macular pathology.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700