Reducing the radiation dose for low-dose CT of the paranasal sinuses using iterative reconstruction: Feasibility and image quality
详细信息    查看全文
文摘

Purpose

To evaluate image quality of dose-reduced CT of the paranasal-sinus using an iterative reconstruction technique.

Methods

In this study 80 patients (mean age: 46.9 ¡À 18 years) underwent CT of the paranasalsinus (Siemens Definition, Forchheim, Germany), with either standard settings (A: 120 kV, 60 mAs) reconstructed with conventional filtered back projection (FBP) or with tube current-time product lowering of 20 % , 40 % and 60 % (B: 48 mAs, C: 36 mAs and D: 24 mAs) using iterative reconstruction (n = 20 each). Subjective image quality was independently assessed by four blinded observers using a semiquantitative five-point grading scale (1 = poor, 5 = excellent). Effective dose was calculated from the dose-length product. Mann-Whitney-U-test was used for statistical analysis.

Results

Mean effective dose was 0.28 ¡À 0.03 mSv(A), 0.23 ¡À 0.02 mSv(B), 0.17 ¡À 0.02 mSv(C) and 0.11 ¡À 0.01 mSv(D) resulting in a maximum dose reduction of 60 % with iterative reconstruction technique as compared to the standard low-dose CT. Best image quality was observed at 48 mAs (mean 4.8; p < 0.05), whereas standard low-dose CT (A) and maximum dose reduced scans (D) showed no significant difference in subjective image quality (mean 4.37 (A) and 4.31 (B); p = 0.72). Interobserver agreement was excellent (¦Ê values 0.79-0.93).

Conclusion

As compared to filtered back projection, the iterative reconstruction technique allows for significant dose reduction of up to 60 % for paranasal-sinus CT without impairing the diagnostic image quality.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700