Bent versus straight tips in micropulsed longitudinal phacoemulsification
详细信息    查看全文
文摘
The aim of this study was to evaluate bent and straight phacoemulsification tips to determine which tip is more efficient in removal of lens fragments, using micropulsed longitudinal ultrasound in phacoemulsification.

Design

In vitro laboratory study.

Methods

The John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, was the study setting. Pig lenses hardened in a manner comparable with dense human cataracts were cut into 2-mm cubes and removed with micropulsed longitudinal ultrasound using settings previously shown to be optimally efficient (6 milliseconds on and 6 milliseconds off for a bent tip). To verify this time as most efficient for a straight tip, we also tested times of 5, 6, and 7 milliseconds time on and off. The tips were either straight or with a 20-degree bend. Twenty cubes were used for each comparative run.

Results

For the straight tip, 6 milliseconds on (1.56 ± 0.815 seconds) was significantly more efficient than 7 milliseconds on (2.45 ± 1.56 seconds, p = 0.001) and not significantly more efficient than 5 milliseconds on (1.69 ± 0.86 seconds, p = 0.43). Five milliseconds off time (1.45 ± 0.76s) was more efficient than 6 milliseconds (2.06 ± 1.37 seconds, p = 0.004) and 7 milliseconds off (2.18 ± 1.24s, p = 0.001). The straight tip was more efficient than the bent tip (1.38 ± 0.83 versus 2.93 ± 2.14 seconds, p = 0.006).

Conclusions

Results are contrary to accepted common belief. Micropulsed longitudinal phacoemulsification is more efficient with a straight rather than a bent tip.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700