Hepatic Microwave Ablation Zone Size: Correlation with Total Energy, Net Energy, and Manufacturer-Provided Chart Predictions
详细信息    查看全文
文摘
To determine whether total energy (TE) reaching the microwave (MW) applicator or net energy (NE) exiting the applicator (after correcting for reflectivity) correlates better with hepatic MW ablation zone dimensions than manufacturer-provided chart predictions.

Materials and Methods

Single-applicator, nonoverlapping ablations of 93 liver tumors (0.7–5.9 cm) were performed in 52 adult patients. TE and NE were recorded for each ablation. Long axis diameter (LAD), short axis diameter (SAD), and volume (V) of each ablation zone were measured on magnetic resonance imaging or computed tomography after the procedure and retrospectively compared with TE; NE; and manufacturer-provided chart predictions of LAD, SAD, and V using correlation and regression analyses.

Results

For treated tumors, mean (± SD) TE and NE were 49.8 kJ (± 22.7) and 36.4 kJ (± 19.4). Mean LAD, SAD, and V were 5.8 cm (± 1.3), 3.7 cm (± 0.8), and 44.1 cm3 (± 25.4). Correlation coefficients (95% confidence interval) with LAD, SAD, and V were 0.46 (0.28, 0.61), 0.52 (0.36, 0.66), and 0.52 (0.36, 0.66) for TE; 0.42 (0.24, 0.58), 0.55 (0.39, 0.68), and 0.53 (0.36, 0.66) for NE; and 0.51 (0.34, 0.65), 0.63 (0.49, 0.74), and 0.60 (0.45, 0.73) for chart predictions. Using regression analysis and controlling for TE, SAD was 0.34 cm larger in patients with cirrhosis than in patients without cirrhosis.

Conclusions

Correcting for reflectivity did not substantially improve correlation of energy values with MW ablation zone size parameters and did not outperform manufacturer-provided chart predictions. Correlations were moderate and variable using all methods. The results suggest a disproportionate influence of tissue factors on MW ablation results.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700