On-line extractive separation in flow injection analysis based on polymer inclusion membranes: A study on membrane stability and approaches for improving membrane permeability
详细信息    查看全文
文摘
The effect of temperature on the sensitivity and sampling rate is studied for a flow injection analysis (FIA) system that uses a membrane separation cell fitted with a polymer inclusion membrane (PIM) for the determination of Zn(II). A temperature of 50 ¡ãC for the flowing donor and acceptor solutions and the membrane separation cell improves the sensitivity and the sampling rate relative to 20 ¡ãC up to 10-fold and 2-fold, respectively.

Studies on the stability of the PIM are reported that show a limited loss of the membrane liquid phase into the aqueous phases used in the FIA system but this has exhibited a negligible effect on the amount of Zn(II) transported across the membrane.

Most importantly, the extent of leaching of the PIM components is shown to depend on the nature of the aqueous phase with the membrane eventually reaching a stable composition.

It is also shown that the application of ultrasound to the membrane separation cell leads to a slight increase in sensitivity without affecting the long term membrane stability.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700