Understanding the effect of noise on electrical stimulation sequences in cochlear implants and its impact on speech intelligibility
详细信息    查看全文
文摘
The present study investigates the most important factors that limit the intelligibility of the cochlear implant (CI) processed speech in noisy environments. The electrical stimulation sequences provided in CIs are affected by the noise in the following three manners. First of all, the natural gaps in the speech are filled, which distorts the low-frequency ON/OFF modulations of the speech signal. Secondly, speech envelopes are distorted to include modulations of both speech and noise. Lastly, the N-of-M type of speech coding strategies may select the noise dominated channels instead of the dominant speech channels at low signal-to-noise ratio's (SNRs). Different stimulation sequences are tested with CI subjects to study how these three noise effects individually limit the intelligibility of the CI processed speech. Tests are also conducted with normal hearing (NH) subjects using vocoded speech to identify any significant differences in the noise reduction requirements and speech distortion limitations between the two subject groups. Results indicate that compared to NH subjects CI subjects can tolerate significantly lower levels of steady state speech shaped noise in the speech gaps but at the same time can tolerate comparable levels of distortions in the speech segments. Furthermore, modulations in the stimulus current level have no effect on speech intelligibility as long as the channel selection remains ideal. Finally, wrong maxima selection together with the introduction of noise in the speech gaps significantly degrades the intelligibility. At low SNRs wrong maxima selection introduces interruptions in the speech and makes it difficult to fuse noisy and interrupted speech signals into a coherent speech stream.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700