文摘
Impedance spectroscopy transduction combined with the immunosensor technology has been used for the determination of atrazine, a herbicide The sensor electrode was based on the immobilization of anti-atrazine antibody by affinity binding onto a polypyrrole film N-substituted by nitrilotriacetic acid NTA electrogenerated on a gold electrode. The poly NTA film was previously modified by the coordination of Cu2 ions by the chelating NTA centers. The anti-atrazine antibody Fab fragment K47 modified with histidine-tag was then anchored by affinity interactions between the histidine-tag and the coordinated Cu2. Cyclic voltammetry experiments confirm that the antibody immobilization and the resulting immunosensor were applied to the impedimetric detection of atrazine without reagent and labelling step. The immunoreaction of atrazine on the attached anti-atrazine antibody directly triggers an increase in the charge transfer resistance proportional to the atrazine concentration, allowing the detection of extremely low atrazine concentration, namely 10 pg mL−1.