文摘
A total of 91 patients with 112 lesions received 2.25-mm sirolimus-eluting stents (SESs), and these lesions were compared with those treated with SESs of ≥2.5-mm diameter in the same procedure (n = 109). The reference diameters were 1.88 ± 0.34 and 2.52 ± 0.57 mm, respectively (p <0.01). At follow-up, the late lumen loss was 0.07 ± 0.48 mm for the 2.25-mm SES versus 0.03 ± 0.38 mm for the larger SES (p = 0.5), and the binary restenosis rate was 10.7 % versus 3.9 % , respectively (p = 0.1). The 12-month target lesion revascularization rate was 5.5 % . In conclusion, 2.25-mm SESs were associated with low rates of clinical and angiographic late complications.