Cytomegalovirus quantification in plasma by an automated real-time PCR assay
详细信息    查看全文
文摘
Background

Sensitive quantitation of cytomegalovirus (CMV) DNA in blood is helpful for the diagnosis of CMV infection or reactivation and the monitoring of transplanted patients.

Objectives

We compared a new PCR assay coupled with an automated extraction system (CMV real-time PCR, Abbott Molecular, Des Plaines, IL, USA) to a previously validated method (ultrasensitive Cobas Amplicor CMV DNA Monitor, Roche Molecular, Indianapolis, IN, USA).

Results

Using limiting dilutions of CMV DNA positive plasma, the two assays had a similar detection threshold ranging between 20 and 45 copies/ml. Coefficients of variation of CMV real-time PCR assay varied from 1 to 12 % for CMV DNA levels between 10,000 and 20 copies/ml. Viral loads assessed by the two methods on 179 clinical samples showed an overall concordance of 89 % and an excellent correlation (R = 0.94). Discrepancies were only observed for samples with low CMV DNA levels (<300 copies/ml); 18 samples were positive by CMV real-time PCR only, and 2 samples by ultrasensitive Cobas CMV only. Values obtained by CMV real-time PCR were on average 0.4 log higher than those of ultrasensitive Cobas CMV. Successive samples of transplanted patients with evidence of CMV infection/reactivation revealed that CMV real-time PCR assay was positive earlier and for a longer period of time after treatment initiation.

Conclusions

Although both assays had similar analytical performances, the CMV real-time PCR assay has the advantages of automated extraction and higher dynamic range, and shows a trend for an improved sensitivity that might impact on clinical decisions.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700