Anterior capsule opacification after femtosecond laser-assisted cataract surgery: Clinical classification versus Scheimpflug device densitometry values
详细信息    查看全文
文摘
To compare the clinical classification of anterior capsule opacification (ACO) after femtosecond laser–assisted cataract surgery with the mean density values of ACO provided by rotating Scheimpflug device (Pentacam HR) densitometry software and to determine which densitometry method correlates best with the clinical classification.

Setting

Ophthalmology Department, Donostia University Hospital, Donostia-San Sebastian, Spain.

Design

Prospective comparative study.

Methods

Femtosecond laser–assisted cataract surgery was performed using the Victus platform between June 2014 and March 2015. Inclusion criteria were age between 55 years and 85 years, a pupil diameter larger than 6.0 mm in full mydriasis, no intraoperative complications, a curvilinear anterior capsulotomy without tears, and an intraocular lens in the correct intracapsular position at the end surgery. The ACO was measured by a clinical classification ranging from 0 to 4. In addition, ACO density was measured with the Scheimpflug device using 3 densitometry methods (area, linear, and peak).

Results

The study comprised 32 eyes of 32 patients. Area and linear densitometry values provided by the Scheimpflug device had a strong correlation with the values obtained by clinical classification, whereas peak densitometry values had a very weak correlation at 6 months (area densitometry: Spearman ρ = 0.78; P < .0005; linear densitometry: ρ = 0.73; P < .0005; peak densitometry ρ = 0.21; P = .2).

Conclusion

The Scheimpflug device provided an objective measurement of ACO after cataract surgery.