Increased detection of smear-negative pulmonary tuberculosis by GeneXpert MTB/RIF® assay after bleach concentration
详细信息    查看全文
文摘
The GeneXpert MTB/RIF assay (Xpert) was endorsed as the initial diagnostic tool in people suspected of human immunodeficiency virus-associated or drug-resistant tuberculosis (TB). However, information regarding the performance of Xpert for diagnosing smear-negative TB in high burden settings remains limited. We evaluated the diagnostic accuracy of Xpert and the impact of bleach concentration on the performance of Xpert using smear-negative sputum samples from human immunodeficiency virus-negative patients.

Methods

One spot and one morning smear-negative sputum samples per patient were examined using Xpert and culture at the Mycobacteriology Research Center of Jimma University, Ethiopia. The sputum culture on both Löwenstein–Jensen and/or Mycobacteria Growth Indicator Tube was the gold-standard.

Results

Of 185 smear-negative presumptive pulmonary TB cases, 19 (10.3%) had culture-proven TB. The sensitivity of Xpert on spot and morning sputum was similar (63.2%). Testing two specimens per patient insignificantly increased the sensitivity of Xpert. Bleach concentration and pelleting improved the sensitivity of Xpert over unprocessed sputum in paired samples (73.8% vs. 63.2%) without affecting the specificity (95%). Bleach concentration and pelleting allowed an additional seven cases of TB (missed on the first and second direct Xperts) to be detected, five of which were from culture-negative cases.

Conclusion

Testing of a single sputum sample by Xpert can reach reasonable sensitivity and results would be available on the same day, avoiding loss of patients and treatment delay. The sensitivity of Xpert was improved after bleach concentration and pelleting, although its added value needs further study on a larger scale.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700