Flexural Rigidity, Taper Angle, and Contact Length Affect Fretting of the Femoral Stem Trunnion in Total Hip Arthroplasty
详细信息    查看全文
文摘
Modularity at the head-neck junction in total hip arthroplasty allows for intraoperative adjustments but may be a source of metallic debris. We determined how flexural rigidity, taper angle, contact length, and lever arm affect fretting and corrosion at this junction.

Methods

A total of 77 metal-on-polyethylene total hip arthroplasties retrieved over a 10-year period at a single institution were obtained. Head tapers and stem trunnions were graded for fretting and corrosion.

Results

Stem fretting was inversely related to rigidity and taper angle, while positively correlated to contact length. Head fretting and head and stem corrosion were not associated with any of these parameters.

Conclusion

Design and assembly factors at the modular head-neck connection affected stem fretting among the retrieved components, suggesting that these parameters are important to consider when choosing a modular system.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700