Fast readout of GEM detectors for medical imaging
详细信息    查看全文
文摘
We describe the design and implementation of a fast data acquisition (DAQ) system for Gas Electron Multiplier (GEM) trackers applied to imaging and dosimetry in hadrontherapy. Within the AQUA project of the TERA foundation a prototype of Proton Range Radiography of 30¡Á30 cm2 active area has been designed and built to provide in-beam integrated density images of the patient before treatment. It makes use of a pair of GEMs to record position and direction of protons emerging from the target.

A fast data acquisition rate close to 1 MHz will allow obtaining a good resolution in-beam proton radiography in a few seconds. A dedicated fast front-end circuit for GEM detectors (GEMROC by AGH-Crakow University) is read by the FPGA based DAQ card (GR_DAQ), developed by the AQUA group. The same system is under evaluation (within the ENVISION European project) to realize the in-vivo dosimetry, based on detecting secondary light particles during the treatment of the patient.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700