CT scans were acquired with an active breathing control device at multiple breath-hold states. Three inverse optimized plans were generated for eight left-sided breast cancer patients: one static IMRT plan optimized at end exhale, two (MIGA) plans based on a MIGA representation of normal breathing, and a MIGA representation of deep breathing, respectively. Breast and nodal targets were prescribed 52.2 Gy, and a simultaneous tumor bed boost was prescribed 60 Gy.
With normal breathing, doses to the targets, heart, and left anterior descending (LAD) artery were equivalent whether optimizing with MIGA or on a static data set. When simulating motion due to deep breathing, optimization with MIGA appears to yield superior tumor-bed coverage, decreased LAD mean dose, and maximum heart and LAD dose compared with optimization on a static representation.
For left-sided breast-cancer patients, inverse-based optimization accounting for motion due to normal breathing may be similar to optimization on a static data set. However, some patients may benefit from accounting for deep breathing with MIGA with improvements in tumor-bed coverage and dose to critical structures.