Analysis of Online and Offline Head and Neck Image-guided Radiation Therapy
详细信息    查看全文
文摘
The introduction of a daily image-guided radiation therapy (IGRT) program is an important step. It has implications for the radiation therapy team in terms of accuracy, workflow, and decision making. This study assesses how successful a radiation therapy department has been in using this technology and the accuracy of individual decision making when comparing online and offline image match data.

Methods

Twenty intensity-modulated radiation therapy head and neck patients had their IGRT data assessed retrospectively. Online analysis was completed based on a 0-mm action threshold. Offline analysis was then conducted on the same data. Any discrepancies in decision making were then assessed.

Results

Results indicated that the treating radiation therapy team was able to image match consistently when benchmarked against their colleagues in the offline environment. Analysis of online versus offline corrections in each of the three orthogonal directions showed strong agreement. Further analysis revealed no statistically significant differences for systematic errors, whereas a statistically significant but small difference was present for random error.

Conclusion

In this age of sophisticated equipment, daily IGRT is a valuable modality. However, the introduction of daily online IGRT inclusive of a 0-mm action threshold for head and neck IGRT requires careful consideration and evidence that such accuracy can be achieved. Ultimately, it is still the radiation therapist who must make the decision, which places great importance on the competency of the treating radiation therapy team.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700