Dosimetric Impact of Interfraction Catheter Movement in High-Dose Rate Prostate Brachytherapy
详细信息    查看全文
文摘

Purpose

To evaluate the impact of interfraction catheter movement on dosimetry in prostate high-dose-rate (HDR) brachytherapy.

Methods and Materials

Fifteen patients were treated with fractionated HDR brachytherapy. Implants were performed on day 1 under transrectal ultrasound guidance. A computed tomography (CT) scan was performed. Inverse planning simulated annealing was used for treatment planning. The first fraction was delivered on day 1. A cone beam CT (CBCT) was performed on day 2 before the second fraction was given. A fusion of the CBCT and CT was performed using intraprostatic gold markers as landmarks. Initial prostate and urethra contours were transferred to the CBCT images. Bladder and rectum contours were drawn, and catheters were digitized on the CBCT. The planned treatment was applied to the CBCT dataset, and dosimetry was analyzed and compared to the initial dose distribution. This process was repeated after a reoptimization was performed, using the same constraints used on day 1.

Results

Mean interfraction catheter displacement was 5.1 mm. When we used the initial plan on day 2, the mean prostate V100 (volume receiving 100 Gy or more) decreased from 93.8 % to 76.2 % (p < 0.01). Rectal V75 went from 0.75 cm3 to 1.49 cm3 (p < 0.01). A reoptimization resulted in a mean prostate V100 of 88.1 % , closer to the initial plan (p = 0.05). Mean rectal V75 was also improved with a value of 0.59 cm3. There was no significant change in bladder and urethra dose on day 2.

Conclusions

A mean interfraction catheter displacement of 5.1 mm results in a significant decrease in prostate V100 and an increase in rectum dose. A reoptimization before the second treatment improves dose distribution.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700