Ex vivo pulmonary nodule detection with miniaturized ultrasound convex probes
详细信息    查看全文
文摘
The intraoperative localization of small and deep pulmonary nodules is often difficult during minimally invasive thoracic surgery. We compared the performance of three miniaturized ultrasound (US) convex probes, one of which is currently used for thoracic endoscopic diagnostic procedures, for the detection of lung nodules in an ex vivo lung perfusion model.

Methods

Three porcine cardiopulmonary blocks were perfused, preserved at 4°C for 6 h and reconditioned. Lungs were randomly seeded with different patterns of echogenicity target nodules (9 water balls, 10 fat, and 11 muscles; total n = 30). Three micro-convex US probes were assessed in an open setting on the pleural surface: PROBE 1, endobronchial US 5–10 MHz; PROBE 2, laparoscopic 4–13 MHz; PROBE 3, fingertip micro-convex probe 5–10 MHz. US probes were evaluated regarding the number of nodules localized/not localized, the correlation between US and open specimen measurements, and imaging quality.

Results

For detecting target nodules, the sensitivity was 100% for PROBE 1, 86.6% for PROBE 2, and 78.1% for PROBE 3. A closer correlation between US and open specimen measurements of target diameter (r = 0.87; P = 0.0001) and intrapulmonary depth (r = 0.97; P = 0.0001) was calculated for PROBE 1 than for PROBES 2 and 3. The imaging quality was significantly higher for PROBE 1 than for PROBES 2 and 3 (P < 0.04).

Conclusions

US examination with micro-convex probes to detect pulmonary nodules is feasible in an ex vivo lung perfusion model. PROBE 1 achieved the best performance. Clinical research with the endobronchial US micro-convex probe during minimally invasive thoracic surgery is advisable.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700