文摘
The dose sensitivity of 4 radiochromic films: EBT3, MD-V3, HD-V2 and HD-810. Median dose using a flatbed scanner and red pixels: 1.4, 11.1, 64 and 61 Gy. Dose rate independence within ±6% from 0.2 to 30 Gy/min. MD-V3 film most compatible with dose range of "3D" radio-fluorogenic (RFG) gel. Start-stop delay times of X-ray and γ-ray sources measured.