Does cancellous bone compaction due to insertion of a blade implant influence the cut-out resistance? A biomechanical study
详细信息    查看全文
文摘

Background

For the treatment of hip fractures helically shaped implants, like the Dynamic Hip Screw (DHS) Blade, are often used. One consequence of blade implantation, the compaction of cancellous bone, is still believed to increase cut-out resistance. This in vitro study investigates implant anchorage of Dynamic Hip Screw Blades in femoral heads due to insertion with or without predrilling under cyclic physiological loading conditions.

Methods

Six pairs of fresh frozen (− 20 °C) human cadaveric proximal femora were instrumented with DHS Blades. Bone pairs were randomly assigned to two study groups: 1) predrilled; 2) non-predrilled. Prior instrumentation, bone mineral density was determined in the center of the femoral head by Xtreme-CT measurement. After instrumentation biomechanical testing was performed under cyclic loading. The bone-implant interface was monitored by means of fluoroscopic imaging throughout the experiment. Paired t-tests were performed to identify differences regarding bone mineral density, stiffness and cycles to failure.

Findings

No significant differences were found between study groups with regard to axial stiffness (P = 0.626) and number of cycles to failure (P = 0.961).

Interpretation

This in vitro study did not show differences in biomechanical stability of proximal femora instrumented with a helical blade implant with or without predrilling. Clinically, the findings suggest that predrilling may be performed to ease the surgical procedure without compromising the implant anchorage.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700