This paper presents the results of an investigation on polarity effects in total-body irradiation (TBI) dosimetry. Thimble (NE2571, 0.6 cc) and plane-parallel (Markus NE2534 0.055 cc) chambers were investigated in a 30 × 30 × 30-cm3 acrylic phantom in TBI conditions (6-MV x-rays). The thimble chamber was positioned at the midline and at the entrance and exit Dmax (1.5 cm from the phantom surface) levels. The Markus chamber, which is generally used for skin dose estimations, was positioned at various depths from the entrance surface of the phantom (from 0- to 2-cm depth). The polarity factor (Ppol) was defined as (Q+ + Q−)/2Q−, where Q+ and Q− were the collected charges at positive and negative bias voltage, respectively. The variations of Ppol with many parameters (absorbed dose, dose rate, the presence or absence of a 1-cm acrylic spoiler, irradiated cable length) were investigated. Results show that Ppol is quite small (within 1.002 for on-axis measurements and 1.005 for off-axis measurements) for the NE2571 chamber when the beam spoiler is placed. Ppol was significantly higher without the beam spoiler (within 1.008 for on-axis measurements, up to 1.02 for off-axis measurements). Concerning the Markus chamber, for on-axis skin dose measurements, Ppol was found to be less than unity (around 0.988) or more than unity (around 1.0035), respectively, with and without the beam spoiler. Possible “directional effects” of the currents generated in the cable were investigated for both chambers and found to be insignificant. This shows that the application of Ppol correction has to be considered a reliable procedure in minimizing these effects. When the beam spoiler is placed, the cable has to be drawn to minimize the portion of cable just outside the beam; if this is not the case, Ppol may significantly vary (for the NE2571 chamber values up to 1.0035 were found for on-axis measurements).