Two experts and four intermediate-level echocardiographers (nonexperts) retrospectively performed analysis of three-dimensional transesophageal echocardiographic images to generate topographic mitral valve models in 90 patients with degenerative MVP. All echocardiographers performed both AIUS and manual segmentation in sequential weekly sessions. The results were compared with surgical findings.
Manual segmentation by nonexperts had significantly lower sensitivity (60% vs 90%, P < .001), specificity (91% vs 97%, P = .001), and accuracy (83% vs 95%, P < .001) compared with experts. AIUS significantly improved the accuracy of nonexperts (from 83% to 89%, P = .003), particularly for lesions involving the A3 (from 81% to 94%, P = .006) and P1 (from 78% to 88%, P = .001) segments, presumably related to anatomic variants of the annulus that made tracking more challenging. AIUS required significantly less time for image analysis by both experts (1.9 ± 0.7 vs 9.9 ± 3.5 min, P < .0001) and nonexperts (5.0 ± 0.5 vs 13 ± 1.5 min, P < .0001), especially for complex lesions.
Anatomic assessment of mitral valve pathology by three-dimensional transesophageal echocardiography is experience dependent. A semiautomated algorithm using AIUS improves accuracy and efficiency in localizing MVP by less experienced operators.