3D microfluidic origami electrochemiluminescence immunodevice for sensitive point-of-care testing of carcinoma antigen 125
详细信息    查看全文
文摘
In this paper, a 3D microfluidic origami device combined with electrochemiluminescence (ECL) immunosensor was introduced for sensitive point-of-care testing of carcinoma antigen 125 in clinical serum samples for the first time. This microfluidic origami device was fabricated by directly screen-printing carbon working, counter electrodes and Ag/AgCl reference electrode including their conductive pads on wax-patterned pure cellulose paper, and was activated by folding to form a 3D electrochemical cell. To construct a sensitivity-enhanced sandwich-type ECL immunosensor, gold nanoparticles (AuNPs) was synthesized to modify the working electrode to provide a good pathway of electron transfer and enhance the immobilized amount of capture antibody (McAb1), and luminol functionalized gold nanoparticles (Lu-AuNPs) was synthesized to label signal antibody (McAb2). With the aid of a facile device-holder, cyclic voltammetry was applied on microfluidic origami ECL immunodevice to trigger ECL. Under optimal conditions, this microfluidic origami ECL immunodevice provided a good linear response range from 0.01 to 100 U mL?1 with a detection limit of 0.0074 U mL?1. The proposed ECL immunodevice presented high sensitivity and stability, and could be applied in point-of-care testing of other tumor markers for remote regions and developing countries.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700